Variable Area Flowmeters

Variable area flowmeters consist of a frame, tapered metering tube, and a float. They are designed to indicate rates of flow for various liquids and gases.

How They Work
Fluid enters the inlet of the metering tube exerting a force on the float. As the force overcomes the weight of the float, the float rises in the tube. When the forces of gravity, float weight, and the fluid force reach an equilibrium, the float stabilizes in the tube. The position of the float at equilibrium corresponds directly to the flow rate. The flow rate is indicated on a scale etched directly onto the metering tube. This scale may be either a direct reading scale or a reference scale. A reference scale requires the use of a calibration data chart.

Variable area flowmeters must be mounted vertically with the flow moving upward through the metering section to assure accuracy. Within this restriction, they offer the flexibility of being bench-mounted or mounted directly in a pipeline, control panel, bypass system, or backup system for electronic controllers.

Description
Aneer Engineer’s flowmeters feature an easy-to-read scale, permanently etched onto a heavy glass flow tube. This scale permits measurement of a wide range of gases and fluids under a variety of pressure and temperature conditions.
Each flowmeter is equipped with two, panel-mount retaining nuts for easy industrial panel mounting or with an optional tripod base for easy bench setup in laboratory applications. Pipe-to-tubing or pipe-to-hose adaptors are also available.

150 mm Flowmeter Tube
Because of its superior resolution, this flowmeter tube is recommended for laboratory applications. Accuracy is ±2% scale with ±.5% repeatability. The elongated scale permits critical flows to be gauged more accurately and read more easily. Conversion charts for air, argon, carbon dioxide, helium, hydrogen, nitrogen, and oxygen are supplied with each unit.

All flowmeters are shipped fully assembled.

Specifications

Maximum Operating Pressure 200 psig (Teflon units - 60 psig)
Maximum Media Temperature

2500F (Teflon units - 1500F)

Inlet/Outlet Connections 1/4" FNPT/Suitable for Tubes

Glass Tubes
Glass tube variable area flowmeters are economical and allow the user to see float movement. They offer the greatest variety of styles, sizes, and materials of construction and also offer tube interchangeability.

Materials of Construction
Materials of construction are an important consideration. End fitting materials are available in brass, stainless steel, and Teflon material. Brass units may be used with inert and noncorrosive media. Stainless steel and Teflon units are recommended for nontoxic, corrosive service applications.

Gas Conversion Factors
To find flow for gases other than air, multiply the flow rate values for air by the applicable conversion factor.

Gas Factor*
Acetylene 1.04
Ammonia 1.30
Argon 0.85
Butane 0.70
Carbon Dioxide 0.81
Carbon Monoxide 1.01
Ethane 0.98
Ethylene 1.02
Helium 2.69
Hydrogen 3.81
Methane 1.35
Neon 1.20
Nitrogen 1.01
Oxygen 0.95
Propane 0.80

*Approximately equal to the square root of the reciprocal of the specific gravity of the intended gas.

Go Top
Home High Purity Gas Distribution System Accessories Medical Equipment Cutting & Welding Chromatography Supply Query Product Animation Material Compatibility Conversion Chart Glossary of Terms Abbreviations & Symbols Technical Information Corporate Information Upcoming Events Catalogue Request ISO Certification